FDA Class I Medical Devices

Nova Biomedical Corporation: StatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.

Nova Biomedical Corporation · Affected quantity: 8,939 units

Hazard
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scann…

Details
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the…

Products affected
StatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.

Recall status
Ongoing

Reference: FDA Z-3131-2026 · Issued September 23, 2026 · Class I · Indexed by RecallSweep from the official FDA feed.