Datex-Ohmeda, Inc. recalls

RecallSweep is tracking 4 recall notices naming Datex-Ohmeda, Inc. in the last 180 days, issued by FDA. The most recent is dated September 23, 2026. They cover Medical Devices.

This page lists what the issuing agencies published. It is not a statement about Datex-Ohmeda, Inc. beyond those notices — open each official notice for the authoritative detail.

Recall notices naming Datex-Ohmeda, Inc.

4 recalls.

FDA Class II Medical Devices

Datex-Ohmeda, Inc.: Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

Affected quantity: 7 units

Hazard
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alar…

Details
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect. Distribution: Worldwide - US Nationwide distribution.. Recall #Z-3191-2026, initiated 20260730. Voluntary: Firm initiated.

Products affected
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

Recall status
Ongoing

FDA Class II Medical Devices

Datex-Ohmeda, Inc.: Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field repla…

Affected quantity: 410 units

Hazard
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alar…

Details
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect. Distribution: Worldwide - US Nationwide distribution.. Recall #Z-3192-2026, initiated 20260730. Voluntary: Firm initiated.

Products affected
Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665

Recall status
Ongoing

FDA Class II Medical Devices

Datex-Ohmeda, Inc.: Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000

Affected quantity: 140 units

Hazard
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alar…

Details
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect. Distribution: Worldwide - US Nationwide distribution.. Recall #Z-3190-2026, initiated 20260730. Voluntary: Firm initiated.

Products affected
Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000

Recall status
Ongoing

FDA Class II Medical Devices

Datex-Ohmeda, Inc.: Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000

Affected quantity: 475 units

Hazard
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alar…

Details
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect. Distribution: Worldwide - US Nationwide distribution.. Recall #Z-3189-2026, initiated 20260730. Voluntary: Firm initiated.

Products affected
Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000

Recall status
Ongoing