Datex-Ohmeda, Inc.: Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field repla…

Datex-Ohmeda, Inc. · Affected quantity: 410 units

Hazard
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alar…

Details
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect. Distribution: Worldwide - US Nationwide distribution.. Recall #Z-3192-2026, initiated 20260730. Voluntary: Firm initiated.

Products affected
Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665

Recall status
Ongoing

Reference: FDA Z-3192-2026 · Issued September 23, 2026 · Class II · Indexed by RecallSweep from the official FDA feed.